This paper describes a simple, inexpensive, highly sensitive, and efficient electrochemical method to determine Estradiol in pharmaceutical formulations. The oxidation reaction on the electrode surface was electrochemically characterized by cyclic voltammetry (CV) and square wave voltammetry (SWV). The investigation of Estradiol at carbon paste electrode revealed a non-reversible oxidation peak at +640 mV vs. Ag/AgCl, which was used for electrochemical detection of Estradiol. The operating parameters (pH, frequency, step potential, and amplitude) were optimized in relation to the peak current intensity, and a calibration curve was set up in a concentration range of 0.059 - 2.997 mg L-1, with a detection limit of 21.85 µg L-1. After calibration curve was plotted, the developed procedure was applied to determine Estradiol in parmacheutical formulation. These results show that the proposed method can be used for Estradiol quantification in pharmaceutical formulations with high sensitivity, specificity, stability, and reproducibility, and can be applied in analytical routines in laboratories of quality control analysis in drugs.
Cookies
We use strictly necessary cookies to run the site and keep you signed in. With your permission we would also use analytics (Google Analytics, Google Tag Manager, Yandex Metrica) to see how the site is used, Google AdSense to show ads, and the Meta Pixel to measure our own ads on Facebook and Instagram. They stay off unless you accept, and you can change your mind at any time. How we use cookies · KVKK notice (Türkiye)