Measurement of sirolimus concentrations in human blood using an automated electrochemiluminescence immunoassay (ECLIA): a multicenter evaluation.

Measurement of sirolimus concentrations in human blood using an automated electrochemiluminescence immunoassay (ECLIA): a multicenter evaluation.

Stove, Veronique;Ramos, Pedro Alía;Wallemacq, Pierre;Vogeser, Michael;Schuetzenmeister, Andre;Schmiedel, Christian;Shipkova, Maria;
Clinical chemistry and laboratory medicine 2018 Vol. 56 pp. 764-775
318
stove2018measurementclinical

Abstract

Therapeutic drug monitoring (TDM) of sirolimus is essential in transplant recipients. We evaluated the performance of a new electrochemiluminescence immunoassay (ECLIA) for measuring sirolimus concentrations in whole blood at five European laboratories.Study assessments included repeatability, intermediate precision and functional sensitivity (concentration at coefficient of variation [CV] of 20%) experiments. Method comparisons with liquid chromatography-tandem mass spectrometry (LC-MS/MS; reference method) and two immunoassays (chemiluminescent microparticle immunoassay [CMIA] and antibody-conjugated magnetic immunoassay [ACMIA]) were performed using native samples from patients with kidney transplants.Imprecision testing CVs were ≤6.4% and ≤10.7% across the sirolimus concentration range for both repeatability and intermediate precision, respectively. The ECLIA showed excellent functional sensitivity: the CV did not reach 20%; the CV at the assay's limit of quantitation (1.5 μg/L) was 7.0%. Agreement between the ECLIA and LC-MS/MS using native kidney samples was close, with weighted Deming regression analysis yielding a slope of 1.05, an intercept of 0.154 μg/L and a Pearson's correlation coefficient (r) of 0.94, while Bland-Altman analysis showed a combined mean bias of 0.41 μg/L (±2 standard deviation [SD], -1.96 to 2.68). The ECLIA also showed good correlation with the two other immunoassays: the CMIA (slope=0.91, intercept=0.112 μg/L and r=0.89) and the ACMIA (slope=0.99, intercept=0.319 μg/L and r=0.97).The ECLIA showed good precision, functional sensitivity and agreement with other methods of sirolimus measurement used in clinical practice, suggesting that the assay is suitable for TDM in transplant recipients and provides an alternative to LC-MS/MS.

Citation

ID: 72774
Ref Key: stove2018measurementclinical
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
72774
Unique Identifier:
10.1515/cclm-2017-0583
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet