Automated electrocardiogram (ECG) interpretations may be erroneous, and lead to erroneous overreads, including for atrial fibrillation (AF). We compared the accuracy of the first version of a new deep neural network 12-Lead ECG algorithm (Cardiologs®) to the conventional Veritas algorithm in interpretation of AF.24,123 consecutive 12-lead ECGs recorded over 6 months were interpreted by 1) the Veritas® algorithm, 2) physicians who overread Veritas® (Veritas® + physician), and 3) Cardiologs® algorithm. We randomly selected 500 out of 858 ECGs with a diagnosis of AF according to either algorithm, then compared the algorithms' interpretations, and Veritas® + physician, with expert interpretation. To assess sensitivity for AF, we analyzed a separate database of 1473 randomly selected ECGs interpreted by both algorithms and by blinded experts.Among the 500 ECGs selected, 399 had a final classification of AF; 101 (20.2%) had ≥1 false positive automated interpretation. Accuracy of Cardiologs® (91.2%; CI: 82.4-94.4) was higher than Veritas® (80.2%; CI: 76.5-83.5) ( < 0.0001), and equal to Veritas® + physician (90.0%, CI:87.1-92.3) ( = 0.12). When Veritas® was incorrect, accuracy of Veritas® + physician was only 62% (CI 52-71); among those ECGs, Cardiologs® accuracy was 90% (CI: 82-94; < 0.0001). The second database had 39 AF cases; sensitivity was 92% vs. 87% ( = 0.46) and specificity was 99.5% vs. 98.7% ( = 0.03) for Cardiologs® and Veritas® respectively.Cardiologs® 12-lead ECG algorithm improves the interpretation of atrial fibrillation.