Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine

Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine

Kelbl, Martin;Kocis, Jan;Vesely, Radek;Florian, Zdenek;Navrat, Tomas;Vosynek, Petr;
asian spine journal 2015 Vol. 9 pp. 503-510
267
kelbl2015biomechanicalasian

Abstract

Study DesignAn in vitro biomechanical study.PurposeTo evaluate the mechanical properties of the spinal segment in the intact, injured, and stabilized state after fixation by an Arcofix implant.Overview of LiteratureSeveral types of thoracolumbar spine injury necessitates anterior instrumentation. The Arcofix plate represents the latest generation of angular stablity systems. The biomechanical properties of these implants have not been sufficiently studied yet.MethodsA total of ten porcine specimens (levels Th12-L3) were prepared. The tests were performed for intact, injured, and implanted specimens. In each state, the specimen was subjected to a tension load of a prescribed force, and subsequently, twisted by a given angle. The force load was 200 N. The torsion load had a deformation character, i.e., the control variable was the twisting angle and the measured variable was the moment of a couple. The amplitude of the load alternating cycle was 3°. Another parameter that was evaluated was the area of the hysteresis loop. The area corresponds to the deformation energy which is dissipated during the cycle.ResultsA statistically significant difference was found between the intact and injured states as well as between the injured and implanted specimens. The statistical evaluation also showed a statistically different value of the hysteresis loop area. In the case of instability, the area decreased to 33% of the physiological value. For the implanted sample, the area increased to 170% of the physiological value.ConclusionsThe Arcofix implant with its parameters appears to be suitable and sufficiently stable for the treatment of the anterior column of the spine.

Citation

ID: 36893
Ref Key: kelbl2015biomechanicalasian
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
36893
Unique Identifier:
f626880b2728e54136b6b4133f7e9893
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet