Accuracy of intraoperative electrophysiological testing in confirming correct cochlear implant electrode positions.
Al Omari, Ahmad;Nuseir, Amjad;Ata, Majid Bani;Khasawneh, Laith;Alhowary, Ala A;Alzoubi, Firas;
cochlear implants international2019pp. 1-7
245
al-omari2019accuracycochlear
Abstract
: To investigate the accuracy of intraoperative electrophysiological studies in detecting incorrectly positioned electrodes in cochlear implant surgery. : A retrospective chart review. : Tertiary referral centre. : In total, 104 consecutive patients with a mean age of 5 years underwent cochlear implant surgery at our centre between January 2012 and December 2013. All patients were implanted with Cochlear Nucleus Freedom implants. : A retrospective study to compare intraoperative neural response telemetry (NRT), impedance and electrode position using Stenver's transorbital plain X-ray view. : Intraoperative electrophysiological tests for patients with Cochlear Nucleus Freedom implants showed 97% sensitivity and 100% specificity compared with postoperative X-ray imaging. : NRT results for the position of cochlear implants were very accurate when checked by X-ray imaging showing that this technique is sufficient in most cases. Stenver's plain X-ray view is needed in complicated cases with abnormal NRT testing or difficult electrode insertion.