estudo comparativo da qualidade de imagem dos modos de aquisição da pet: validação de um protocolo para reduzir a dose de radiação comparative study on imaging quality in pet acquisition modes: validation of a protocol for reducing the radiation dose

estudo comparativo da qualidade de imagem dos modos de aquisição da pet: validação de um protocolo para reduzir a dose de radiação comparative study on imaging quality in pet acquisition modes: validation of a protocol for reducing the radiation dose

;Solange Amorim Nogueira;Henrique Manoel Lederman;Jairo Wagner;Lilian Yuri Yamaga;Marcelo Livorsi da Cunha;Marcelo Buarque de Gusmão Funari
india international conference on information processing, iicip 2016 - proceedings 2009 Vol. 42 pp. 103-107
241
nogueira2009radiologiaestudo

Abstract

OBJETIVO: O presente experimento visa a validar um protocolo de aquisição em 3D na tomografia por emissão de pósitrons, em substituição ao modo 2D, de forma a reduzir a dose de radiação nos pacientes, sem perda da qualidade de imagens. MATERIAIS E MÉTODOS: Foram realizadas 27 simulações em equipamento Discovery ST, nos modos 2D com quatro minutos de aquisição e 3D com dois e quatro minutos. Utilizou-se um simulador do protocolo da National Electrical Manufacturers Association. No interior deste simulador estão inseridas seis esferas com diferentes diâmetros para a determinação da qualidade de imagem. As aquisições foram comparadas por três médicos nucleares, sem que eles identificassem o modo de aquisição. Cada observador atribuiu o valor igual a 1 quando alguma esfera não foi identificada ou valor 2 para esferas visíveis. RESULTADOS: A análise qualitativa pelo kappa generalizado demonstrou que a frequência de esferas visíveis foi maior no modo 3D com quatro minutos (85%) e a porcentagem de concordância também foi maior (88,9%), com kappa generalizado = 0,725 [0,507;0,942]. CONCLUSÃO: O modo 3D com quatro minutos de aquisição e com menores atividades de FDG-18F pode ser utilizado em pacientes com biótipo equivalente ao simulador, sem perda de qualidade de imagem.
OBJECTIVE: The present study is aimed at validating a 3D acquisition protocol for positron emission tomography as a replacement for the 2D mode, to reduce the radiation dose delivered to patients, without any loss in the quality of images. MATERIALS AND METHODS: The study comprised 27 simulations in a Discovery ST equipment with four-minute 2D acquisitions, and two-minute and four-minute 3D acquisitions, utilizing a chest phantom according to the National Electrical Manufacturers Association protocol. Six spheres with different diameters were inserted into this phantom as a means for determining the images quality. The images were blindly reviewed by three experienced nuclear physicians who did not know the acquisition modes. Each observer attributed a score 1 when one of the spheres was not identified, or 2 for visible spheres. RESULTS: The qualitative analysis based on generalized kappa coefficient demonstrated that the frequency of visible spheres was higher with four-minute 3D acquisitions (85%), with an also higher interobserver agreement (88.9%), generalized kappa = 0.725 [0.507;0.942]. CONCLUSION: The protocol with four-minute 3D acquisition with lower 18F-FDG activity can be utilized for patients with a biotype similar to the phantom, without any loss in the imaging quality.

Citation

ID: 158772
Ref Key: nogueira2009radiologiaestudo
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
158772
Unique Identifier:
10.1590/S0100-39842009000200008
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet