Background: VDD pacing system is an alternative to DDD pacing system. Due to different fixation of electrodes in both systems, detected of atrial signal may vary greatly and can be influenced by posture and exercise. Inappropriate atrial sensing leads into loss of atrioventricular synchrony. Haemodynamic changes that occur have a distinctive influence on peripheral levels ANP and BNP.

Methods: Our prospective study encompassed seventy-five patients with implanted VDD or DDD pacing system in year 2000. Using the programmer we directly measured sensing threshold in atrium supine, sitting upright and after standardized exercise. Among measured data we evaluated AV synchrony. Blood was collected from an antecubital vein and plasma proANP and NT-proBNP levels were measured.

Results: Atrial sensing was significantly lower with VDD compared to DDD (p < 0.001). The difference between supine and sitting upright among each pacing system was not significant (p = 0.82). Atrial sensing was significantly lower after exercise compared to supine with VDD (p < 0.05). VDD AV synchrony was lower (p < 0.001). Difference of ANP levels was not significant (p = 0.13). Levels of BNP were lower with VDD (p < 0.05). Levels of ANP, BNP are negatively correlated to AV synchrony ( σ = –0.44, p < 0.001 and σ = –0.54, p < 0.001, respectively). Levels of BNP and ANP are closely correlated ( σ = 0.71, p < 0,001).

Conclusions: Posture did not have any significant effect on atrial sensing. However, in VDD pacing system atrial sensing was affected by exercise. Lower AV synchrony with VDD pacing system suggests lower haemodynamic support. Peripheral levels of natriuretic peptides were associated with AV synchrony. Therefore, haemodynamic effect can be predicted by measuring peripheral levels of either ANP or BNP.

Citation

ID: 158273
Ref Key: iek2005zdravnikiddd
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
158273
Unique Identifier:
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet