Photoacoustic gas monitoring for anesthetic gas pollution measurements and its cross-sensitivity to alcoholic disinfectants.

Photoacoustic gas monitoring for anesthetic gas pollution measurements and its cross-sensitivity to alcoholic disinfectants.

Herzog-Niescery, Jennifer;Steffens, Thomas;Bellgardt, Martin;Breuer-Kaiser, Andreas;Gude, Philipp;Vogelsang, Heike;Weber, Thomas Peter;Seipp, Hans-Martin;
bmc anesthesiology 2019 Vol. 19 pp. 148
319
herzogniescery2019photoacousticbmc

Abstract

Real-time photoacoustic gas monitoring is used for personnel exposure and environmental monitoring, but its accuracy varies when organic solvents such as alcohol contaminate measurements. This is problematic for anesthetic gas measurements in hospitals, because most disinfectants contain alcohol, which could lead to false-high gas concentrations. We investigated the cross-sensitivities of the photoacoustic gas monitor Innova 1412 (AirTech Instruments, LumaSense, Denmark) against alcohols and alcoholic disinfectants while measuring sevoflurane, desflurane and isoflurane in a laboratory and in hospital during surgery.25 mL ethyl alcohol was distributed on a hotplate. An optical filter for isoflurane was used and the gas monitor measured the 'isoflurane' concentration for five minutes with the measuring probe fixed 30 cm above the hotplate. Then, 5 mL isoflurane was added vaporized via an Anesthetic Conserving Device (Sedana Medical, Uppsala, Sweden). After one-hour measurement, 25 mL isopropyl alcohol, N-propanol, and two alcoholic disinfectants were subsequently added, each in combination with 5 mL isoflurane. The same experiment was in turn performed for sevoflurane and desflurane. The practical impact of the cross-sensitivity was investigated on abdominal surgeons who were exposed intraoperatively to sevoflurane. A new approach to overcome the gas monitor's cross-sensitivity is presented.Cross-sensitivity was observed for all alcohols and its strength characteristic for the tested agent. Simultaneous uses of anesthetic gases and alcohols increased the concentrations and the recovery times significantly, especially while sevoflurane was utilized. Intraoperative measurements revealed mean and maximum sevoflurane concentrations of 0.61 ± 0.26 ppm and 15.27 ± 14.62 ppm. We replaced the cross-sensitivity peaks with the 10th percentile baseline of the anesthetic gas concentration. This reduced mean and maximum concentrations significantly by 37% (p < 0.001) and 86% (p < 0.001), respectively.Photoacoustic gas monitoring is useful to detect lowest anesthetic gases concentrations, but cross-sensitivity caused one third falsely high measured mean gas concentration. One possibility to eliminate these peaks is the recovery time-based baseline approach. Caution should be taken while measuring sevoflurane, since marked cross-sensitivity peaks are to be expected.

Citation

ID: 15412
Ref Key: herzogniescery2019photoacousticbmc
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
15412
Unique Identifier:
10.1186/s12871-019-0822-7
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet