Observational, descriptive, exploratory, case study with the objective of measuring the stimulus-response time of the team to alarms monitoring invasive blood pressure (IBP) and analyzing the implications of this time for the safety of the patient. From January to March 2013, 60 hours of structured observation were conducted with registration of the alarms activated by IBP monitors in an adult ICU at a military hospital in the city of Rio de Janeiro. 76 IBP alarms were recorded (1.26 alarms/hour), 21 of which (28%) were attended to and 55 (72%) considered as fatigued. The average response time to the alarms was 2 min. 45 sec. The deficit in human resource and physical layout were factors determining the delay in response to the alarms. The increase in response times to these alarms may compromise the safety of patients with hemodynamic instability, especially in situations such as shock and the use of vasoactive drugs.

Citation

ID: 143709
Ref Key: pergherrevistastimulus-response
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
143709
Unique Identifier:
10.1590/1983-1447.2014.02.43715
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
4/5
Blockchain Upload Locked

Complete all 5 checklist items to tokenize your article

Saymatik Web3.0 Wallet