comparing the corneal curvatures obtained from three different keratometers-iol master, bausch & lomb manual keratometer and topcon kr-8800 autokeratometer

comparing the corneal curvatures obtained from three different keratometers-iol master, bausch & lomb manual keratometer and topcon kr-8800 autokeratometer

;Arpitha Pereira;Ajita Sasidharan
journal of fungi 2018 Vol. 18 pp. 17-20
157
pereira2018guojicomparing

Abstract

AIM: To compare the corneal curvature and to investigate the agreement between three different keratometers.

METHODS: In this prospective study, keratometry was performed using an IOL Master(IM), a Bausch & Lomb manual keratometer(Man)and TOPCON KR-8800 autokeratometer(Top)on 252 eyes of patients recruited from camps for cataract surgery. The average keratometry values were recorded and compared. The agreements between the instruments were analyzed using the Bland-Altman statistical method. The main outcome measure was average keratometry values.

RESULTS:1)IOL Master and Manual keratometer: the mean corneal power was 44.62±1.52 D with the IOL Master and 44.60±1.52 D with the Manual keratometer. The paired t-test demonstrated a statistically significant difference in the mean corneal power between the IOL Master and Manual keratometer(P=0.001). The 95% LOAs of the two devices were -0.22 to 0.22 as shown in the Bland-Altman plot; 2)IOL Master and autokeratometer: the mean corneal power was 44.62±1.52 D with the IOL Master and 44.46±1.53 D with the autokeratometer. The paired t-test demonstrated a statistically significant difference in the mean corneal power between the IOL Master and autokeratometer(P<0.0001). The 95% LOAs of the two devices were -0.24 to 0.55 as shown in the Bland-Altman plot; 3)Autokeratometer and Manual keratometer: the mean corneal power was 44.60±1.52 D with Manual keratometer and 44.46±1.53 D with the autokeratometer. The paired ttest demonstrated a statistically significant difference in the mean corneal power between the autokeratometer and Manual keratometer(P<0.0001). The 95% LOAs of the two devices were -0.30 to 0.57 as shown in the Bland-Altman plot.

CONCLUSION: Keratometry data obtained with different instruments may not be interchangeable, a fact that has important implications for cataract surgeons with respect to both surgical planning and outcomes auditing.

Citation

ID: 135914
Ref Key: pereira2018guojicomparing
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
135914
Unique Identifier:
10.3980/j.issn.1672-5123.2018.1.04
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet