Development of an In-Field Method for the Detection of Barium in Various Water Samples Using Differential Pulse Anodic Stripping Voltammetry
Ridgway, S.;Wajrak, M.;Ridgway, S.;Wajrak, M.;
international journal of electrochemistry2019Vol. 2019
239
s2019developmentinternational
Abstract
This work presents a reliable, cost-effective, rapid and in-field voltammetric method for the detection of barium. The optimized method consists of an ultrathin mercury film deposited in situ on a glassy carbon electrode in dilute potassium chloride without deoxygenation, using differential pulse anodic stripping voltammetry (DP-ASV). Application of the method allowed for the quantitative determination of barium concentration in a variety of waters and brake pad dust samples. Comparative analysis of sample results from DP-ASV with inductively coupled plasma mass spectroscopy (ICP-MS) showed a mean percent difference of 1.8%.