Validation of electronic visual acuity (EVA) measurement against standardised ETDRS charts in patients with visual field loss from inherited retinal degenerations.

Validation of electronic visual acuity (EVA) measurement against standardised ETDRS charts in patients with visual field loss from inherited retinal degenerations.

Jolly, Jasleen K;Juenemann, Kristin;Boagey, Heather;Nadsady, Marie;Bridge, Holly;Maclaren, Robert E;
the british journal of ophthalmology 2019
256
jolly2019validationthe

Abstract

With the increase in clinical trials testing therapy for retinal disease, there is a need to ensure that outcome measures are both accurate and standardised. The US Food and Drug Administration favours the use of visual acuity measured using ETDRS logMAR charts. The loss of visual field can interfere with visual tracking across the charts, leading to increased variability of measurements. Electronic visual acuity (EVA) presents the optotype on the centre of a screen, thereby removing the tracking element of the task, and may provide a more precise measurement.Visual acuity was measured twice using ETDRS charts, EVA automated single letter (E-ETDRS) and EVA single line (EVA-SL) presentation (EMMES). Patients underwent microperimetry (MAIA; Centervue) to determine visual field. We tested 65 patients with rod-cone dystrophies and 41 healthy volunteers.Both participant groups read 2-3 letters more on average on the electronic charts compared with ETDRS. Limits of agreement using a modified Bland-Altman analysis account for replicates were wider in eyes with foveal defects (-9 to 18) compared with eyes without foveal defects (-11 to 15). Electronic charts in the presence of foveal defects reduced the range (-11 to 13).EVA may provide more accurate measures of visual acuity than traditional ETDRS charts in patients when the visual field loss encroached on the central vision. Electronic presentation with a single line of letters was the favoured style reported by patients and should be considered in future interventional clinical trials.

Citation

ID: 57799
Ref Key: jolly2019validationthe
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
57799
Unique Identifier:
bjophthalmol-2019-315124
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet