Quantitative and clinical impact of MRI-based attenuation correction methods in [F]FDG evaluation of dementia.

Quantitative and clinical impact of MRI-based attenuation correction methods in [F]FDG evaluation of dementia.

Øen, Silje Kjærnes;Keil, Thomas Morten;Berntsen, Erik Magnus;Aanerud, Joel Fredrik;Schwarzlmüller, Thomas;Ladefoged, Claes Nøhr;Karlberg, Anna Maria;Eikenes, Live;
ejnmmi research 2019 Vol. 9 pp. 83
230
en2019quantitativeejnmmi

Abstract

Positron emission tomography/magnetic resonance imaging (PET/MRI) is a promising diagnostic imaging tool for the diagnosis of dementia, as PET can add complementary information to the routine imaging examination with MRI. The purpose of this study was to evaluate the influence of MRI-based attenuation correction (MRAC) on diagnostic assessment of dementia with [F]FDG PET. Quantitative differences in both [F]FDG uptake and z-scores were calculated for three clinically available (DixonNoBone, DixonBone, UTE) and two research MRAC methods (UCL, DeepUTE) compared to CT-based AC (CTAC). Furthermore, diagnoses based on visual evaluations were made by three nuclear medicine physicians and one neuroradiologist (PET, PET, PET, PET, PET + MRI, PET + MRI). In addition, pons and cerebellum were compared as reference regions for normalization.The mean absolute difference in z-scores were smallest between MRAC and CTAC with cerebellum as reference region: 0.15 ± 0.11 σ (DeepUTE), 0.15 ± 0.12 σ (UCL), 0.23 ± 0.20 σ (DixonBone), 0.32 ± 0.28 σ (DixonNoBone), and 0.54 ± 0.40 σ (UTE). In the visual evaluation, the diagnoses agreed with PET in 74% (PET), 67% (PET), and 70% (PET) of the patients, while PET + MRI agreed with PET + MRI in 89% of the patients.The MRAC research methods performed close to that of CTAC in the quantitative evaluation of [F]FDG uptake and z-scores. Among the clinically implemented MRAC methods, Dixon should be preferred for diagnostic assessment of dementia with [F]FDG PET/MRI. However, as artifacts occur in Dixon attenuation maps, they must be visually inspected to assure proper quantification.

Citation

ID: 17308
Ref Key: en2019quantitativeejnmmi
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
17308
Unique Identifier:
10.1186/s13550-019-0553-2
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet