Radiotherapy isocenters verified by matching to bony landmarks of the canine and feline head differ when localized using volumetric vs planar imaging.

Radiotherapy isocenters verified by matching to bony landmarks of the canine and feline head differ when localized using volumetric vs planar imaging.

Magestro, Leanne M;Cahoon, Joyce Y;Gieger, Tracy L;Nolan, Michael W;
veterinary and comparative oncology 2019
309
magestro2019radiotherapyveterinary

Abstract

The "gold standard" for verification of patient positioning before linear accelerator-based stereotactic radiation therapy is kilovoltage cone-beam computed tomography (kV-CBCT), which is not uniformly available or utilized; planar imaging is sometimes used instead. The primary aim of this study was to determine if the position of the bony skull, when used as a surrogate for isocenter verification, is different when orthogonal megavoltage portal (MV) or kilovoltage (kV/kV) radiographs are used for image-guidance, rather than kV-CBCT. A secondary aim was to determine the influence of intraobserver variability, body size, and skull conformation on positioning, as determined using these three imaging modalities. Dogs and cats receiving radiotherapy of the head were recruited for this prospective analytical study. Planar (MV portal and kV/kV images) and volumetric (kV-CBCT) images were acquired before treatment, and manually co-registered with reference images. Differences in skull position when matched based on MV portal, kV/kV images, and kV-CBCT were compared. Sixty-five subjects and 148 unique datasets were evaluated. The Wilcoxon rank-sum test was used to evaluate effects of transitioning between imaging modalities. When comparing magnitude of shifts in MV to kV-CBCT, MV to kV/kV, and kV/kV to kV-CBCT, there were statistically significant differences. Results were not measurably impacted by body size, skull conformation, or interobserver differences. Based on shift magnitude and direction, an isotropic setup margin of at least 1 mm should be incorporated within the planning target volume when MV or kV planar imaging is used for position verification. This article is protected by copyright. All rights reserved.

Citation

ID: 4292
Ref Key: magestro2019radiotherapyveterinary
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
4292
Unique Identifier:
10.1111/vco.12522
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet