Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study

Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study

Negro, Roberto W. Dal;Longo, Pietro;Ziani, Orestepaolo Villanis;Bonadiman, Luca;Turco, Paola;
multidisciplinary respiratory medicine 2017 Vol. 12 pp. 1-6
261
negro2017instantmultidisciplinary

Abstract

Abstract Background Inhalation is the preferred route for respiratory drug delivery, but several factors contribute to the variability of the respirable dose fraction. Instant velocity and the dynamic characteristics of the droplet cloud represent crucial factors. Aim was to measure and compare the instant velocity and the consistency of emitted cloud from five different MDIs (A - Salbutamol sulphate 100mcg, GSK; B - Salbutamol sulphate 100mcg, Valeas; C - Salmeterol xinafoate/Fluticasone propionate 25/125mcg, GSK; D - Formoterol fumarate/Bechlomethasone propionate 6/100mcg, Chiesi; E - Formoterol fumarate/Fluticasone dipropionate 5/125mcg, Mundipharma) and one SMI (Tiotropium bromide 5mcg, Boehringer Ingelheim), at different distance from the nozzle and canister filling. Methods Measurements were made at 90, 50, and 10% of canister filling, and at 5, 10, and 20 cm from the nozzle, for a total of 972 puffs. A high speed video photography protocol was adopted and high speed cameras (1.200 frames/sec.) were used. Data were acquired by means of specialized softwares. Temperature, humidity, and vibrations occurrence were strictly controlled during measurements. Statistics: Anova and p < 0.05 were accepted as the minimum significance level. Results MDIs generated different Instant velocities: MDI B generated the highest, while MDI A the lowest. As expected, velocity decreased in proportion to the distance from the nozzle. Except with MDI C, instant velocity decreased significantly over the first 50% of canister emptying, but dropped by >33% at 90% of emptying with all other MDIs (p < 0–037; p < 0.001; p < 0.005, and p < 0.001, respectively). Instant velocity was extremely lower (p < 0.001) and constant for all levels of canister filling (p = ns) with SMI. All MDIs had a very fast jet phase, ranging 0.01–0.03 s at 10 cm, and 0.03–0.05 s at 20 cm from the nozzle, without any significant difference from each other (p = ns). MDIs generated a cloud similarly tight (p = ns) at 10 and 20 cm from the nozzle, while it was extremely wider and constant with the SMI (p = 0–001). Also the cloud turbulence was minimized during the SMI emission. Discussion and Conclusions MDIs are characterized by a substantial variability in both their instant velocity and consistency of the emitted cloud at different levels of canister filling. SMI generates a much slower soft mist cloud which is constantly homogeneous and independent of canister emptying. These peculiarities assessed at bench are suggesting a higher dose consistency and a much more effective therapeutic performance also in real life.

Citation

ID: 19405
Ref Key: negro2017instantmultidisciplinary
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
19405
Unique Identifier:
69874e484a456843f9c6715276f3b5ab
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet