short wave–automated perimetry (swap) versus optical coherence tomography in early detection of glaucoma

short wave–automated perimetry (swap) versus optical coherence tomography in early detection of glaucoma

;Zaky AG;Yassin AT;El Sayid SH
e3s web of conferences 2016 Vol. Volume 10 pp. 1819-1824
179
ag2016clinicalshort

Abstract

Adel Galal Zaky,1 Ahmed Tarek Yassin,2 Saber Hamed El Sayid1 1Ophthalmology Department, Faculty of Medicine, Menoufia University, Shebin El Kom, Menoufia, Egypt; 2Ophthalmology Department, Banha Educational Hospital, Banha, El Kalyobia, Egypt Objective: To assess the role and diagnostic effectiveness of optical coherence tomography (OCT) and short wave–automated perimetry (SWAP) to distinguish between normal, glaucoma suspects, and surely diagnosed glaucomatous eye.Background: Changes in the optic disc and retinal nerve fiber layer (RNFL) often precede the appearance of visual field defect with standard automated perimetry. Unfortunately, RNFL defect can be difficult to identify during clinical examination. Early detection of glaucoma is still controversial, whether by OCT, SWAP, or frequency-doubling technology perimetry.Patients and methods: In this randomized controlled, consecutive, prospective study, a total 70 subjects (140 eyes) were included in the study, divided into three groups: Group A, 10 healthy volunteers (20 eyes); Group B, 30 patients (60 eyes) with glaucoma suspect; and Group C, 30 patients (60 eyes) with already diagnosed glaucomatous eyes.Results: Average RNFL thickness was 75±9.0 in the glaucoma group, 99±15.5 in the control group, and 94±12 in glaucoma suspect. The inferior quadrant was the early parameter affected. There was significant correlation between visual field parameters and RNFL thickness in both glaucoma and glaucoma suspect groups.Conclusion: Both RNFL thickness measured by OCT and SWAP indices are good discrimination tools between glaucomatous, glaucoma suspect, and normal eyes. OCT parameters tend to be more sensitive than SWAP parameters. Keywords: OCT, SWAP, glaucoma, intraocular pressure, RNFL

Keywords

Citation

ID: 159887
Ref Key: ag2016clinicalshort
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
159887
Unique Identifier:
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet