a light-field-based method to adjust on-axis rounded leaf end mlc position to predict off-axis mlc penumbra region dosimetric performance in a radiation therapy planning system

a light-field-based method to adjust on-axis rounded leaf end mlc position to predict off-axis mlc penumbra region dosimetric performance in a radiation therapy planning system

;Jia-Ming Wu;Tsair-Fwu Lee;Shyh-An Yeh;Kuan-Yin Hsiao;Hsin-Hsiung Chen;Pei-Ju Chao;Yi-Ting Chen
spectrochimica acta - part a: molecular and biomolecular spectroscopy 2013 Vol. 2013 pp. -
155
wu2013biomeda

Abstract

Purpose. An analytical and experimental study of split shape dose calculation correction by adjusting the position of the on-axis round leaf end position is presented. We use on-axis corrected results to predict off-axis penumbra region dosimetric performance in an intensity-modulated radiation therapy treatment planning system. Materials and Methods. The precise light-field edge position (Xtang.p) was derived from the on-axis 50% dose position created by using the nominal light field for geometric and mathematical manipulation. Leaf position (Xmlc.p) could be derived from Xtang.p by defining in the treatment planning system for monitor unit calculation. On-axis offset (correction) could be obtained from the position corresponding to 50% of the central axis dose minus the Xmlc.p position. The off-axis 50% dose position can then be derived from the on-axis 50% dose position. Results. The monitor unit calculation of the split shape using the on-axis rounded leaf end MLC penumbra region could provide an under-or overdose of 7.5% per millimeter without an offset correction. When using the on-axis rounded leaf end offset correction to predict the off-axis dose, the difference between the off- and on-axis 50% dose position is within ±1.5 mm. Conclusions. It is possible to achieve a dose calculation within 0.5% error for an adjusted MLC leaf edge location in the treatment planning system with careful measurement and an accurate on-axis offset correction. Dose calculations located at an off-axis spilt shape region should be used carefully due to noncorrectable errors which were found to be up to 10%.

Keywords

Citation

ID: 143487
Ref Key: wu2013biomeda
Use this key to autocite in SciMatic or Thesis Manager

References

Blockchain Verification

Account:
NFT Contract Address:
0x95644003c57E6F55A65596E3D9Eac6813e3566dA
Article ID:
143487
Unique Identifier:
10.1155/2013/461801
Network:
Scimatic Chain (ID: 481)
Loading...
Blockchain Readiness Checklist
Authors
Abstract
Journal Name
Year
Title
5/5
Creates 1,000,000 NFT tokens for this article
Token Features:
  • ERC-1155 Standard NFT
  • 1 Million Supply per Article
  • Transferable via MetaMask
  • Permanent Blockchain Record
Blockchain QR Code
Scan with Saymatik Web3.0 Wallet

Saymatik Web3.0 Wallet